Experimental characterisation of the thermal lesion induced by microwave ablation

TitoloExperimental characterisation of the thermal lesion induced by microwave ablation
Tipo di pubblicazioneArticolo su Rivista peer-reviewed
Anno di Pubblicazione2014
AutoriLopresto, Vanni, Pinto Rosanna, and Cavagnaro M.
RivistaInternational Journal of Hyperthermia
Volume30
Paginazione110-118
ISSN02656736
Parole chiaveAblation Techniques, ablation therapy, adult, animal, animal tissue, Animals, article, Bovinae, Cattle, Electric conductivity, Electrochemical impedance spectroscopy, Heat, Hot Temperature, in vivo study, intestine function, Liver, microwave radiation, microwave therapy, Microwaves, Models, nonhuman, radiation response, sigmoid, Surgery, Temperature dependence, Temperature measurement, Theoretical, theoretical model, thermal injury, treatment planning
Abstract

Purpose: This work focuses on the characterisation of the ablated area induced by a microwave thermal ablation (MTA) procedure. An experimental methodology for establishing a straightforward correlation between the temperature gradient and the changes in the dielectric properties of the tissue is presented and discussed. Materials and methods: Temperature measurements were performed during an ablation procedure in ex vivo bovine liver, at different distances from the antenna, whereas measurements of complex permittivity were conducted in sagittal sections of the ablated samples. The measured temperatures and dielectric properties were then correlated to obtain the dependence of the dielectric properties' spatial variation on the temperature gradient. The obtained correlation has been validated through comparison with previously obtained experimental data. A weighted cubic polynomial function and a weighted sigmoid function have been tested for best-fit interpolation of the measured data. Results: Temperatures in the range 23-105 °C were measured during the MTA procedure, while, after the end of the MTA trials, relative permittivities in the range 7-43 and electric conductivities in the range 0.3-1.8 S/m were measured according to the distance from the antenna's axis. The polynomial function showed better regression coefficients than the sigmoid one for both the relative permittivity (R2= 0.9947 versus R2= 0.9912, respectively) and the conductivity (R2= 0.9919 versus R2= 0.9866, respectively). However, the weighted cubic function showed an unrealistic behaviour for the relative permittivity at temperatures lower than 40 °C. Conclusions: According to the results obtained, information on the changes in the dielectric properties of the tissue under MTA treatment could be inferred from measured temperature data. Once validated by in vivo studies, the proposed methodology could be exploited to develop predictive tools for treatment planning. © 2014 Informa UK Ltd.

Note

cited By 16

URLhttps://www.scopus.com/inward/record.uri?eid=2-s2.0-84896857559&doi=10.3109%2f02656736.2013.879744&partnerID=40&md5=517b559861bdcee5ea64be167bf0b44f
DOI10.3109/02656736.2013.879744
Citation KeyLopresto2014110